Abstract:

Objective

This study evaluates the feasibility and safety of modular Carina surgical robot system for thyroidectomy.

Methods

In this prospective single-arm trial, 8 patients with papillary thyroid carcinoma were enrolled at Yantai Yuhuangding Hospital following pig and cadaver feasibility studies. Carina™ was used to perform robotic thyroidectomy, with outcomes assessing system feasibility and safety.

Results

Between May 2025 and Jul. 2025, all 8 subjects successfully underwent robot-assisted endoscopic thyroid cancer surgery using Carina™: 7 via gasless axillary approach and 1 via bilateral axillo-breast approach (BABA), with no conversions to laparoscopy or open surgery. The average system preparation time was 11.75±5.26 minutes, the average remote operation time was 111.88±25.45 minutes, and the average total operation time was 194.38±27.83 minutes. The system preparation time, remote operation time, and total operation time all showed a stable downward trend. The average physiological load score of the lead surgeon was 1.75±2.92 points, and the average psychological load score was 21.13±10.45 points. Both the physiological load and psychological load scores showed a significant downward trend as the number of surgeries increased. An average of 6.38 ± 2.77 lymph nodes were obtained, and an average of 0.88±1.73 metastatic lymph nodes were present. Postoperative follow-up was conducted for at least 2 month, and no serious complications was observed.

Conclusion

This study demonstrates for the first time that modular robotic thyroidectomy using Carina™ is feasible, safe, and effective.

Key words: Carina? surgical robot system, Robot-assisted, Non-inflated transaxillary approach, Bilateral axillo-breast approach (BABA), Thyroid